Feature TMF10105; AC/DC Power Modules TMF Series, 5 – 30 W • Fully encapsulated power supplies in plasic cas.
Manufacture TRACO POWER
Datasheet
Download TMF10105 Datasheet




TRACO POWER TMF10105
AC/DC Power Modules
TMF Series, 5 – 30 W
Fully encapsulated power supplies in
plasic casing for PCB mount
Certification according to IEC/EN/ES
60601-1 3rd edition for 2xMOPP
Risk management process according to
ISO 14971 including risk
management file
Acceptance criteria for electronic as
semblies according to
IPC-A-610 Level 3
Low leakage current
Operating temperature range:
–25°C to +70°C max.
Protection against short-circuit,
over load and over voltage
Protection class II prepared
5 year product waranty
The TMF Series AC/DC power supply modules are designed and manufactured
based on workmanship standards and risk management to comply with the re-
quirements for quality, reliability and safety of medical equipment. The units are
approved to IEC/EN/ES 60601-1 edition 3.1 for 2 x MOPP (Means Of Patient
Protection) and come along with an ISO 14971 risk management file.
These fully encapsulated modules are for PCB mount. They are designed for
protection class II applications (no earth connection) and feature a low leakage
current (<100 µA). A compact design and excellent EMC considerations facili-
tate the design in. The thermal management enables an operation within a wide
temperature range of -25 to +70°C and the isolation system is designed and ap-
proved for an altitude of 5000 m (AMSL). This makes the power supplies suitable
not only for stationary applications but also for transportable medical equipment.
Models
Order Code
TMF 05105
TMF 05112
TMF 05115
TMF 05124
TMF 10105
TMF 10112
TMF 10115
TMF 10124
TMF 20105
TMF 20112
TMF 20115
TMF 20124
TMF 30105
TMF 30112
TMF 30115
TMF 30124
Output Power
(max.)
5W
5W
5W
5W
10 W
10 W
10 W
10 W
18 W
20 W
20 W
20 W
25 W
30 W
30 W
30 W
Output Voltage
Output Current
(max.)
Efficiency
(typ.)
5.0 VDC
1000 mA
77 %
12 VDC
417 mA
82 %
15 VDC
333 mA
82 %
24 VDC
208 mA
82 %
5.0 VDC
2000 mA
79 %
12 VDC
833 mA
84 %
15 VDC
666 mA
84 %
24 VDC
417 mA
84 %
5.0 VDC
3600 mA
78 %
12 VDC
1667 mA
84 %
15 VDC
1333 mA
84 %
24 VDC
833 mA
84 %
5.0 VDC
5000 mA
82 %
12 VDC
2500 mA
88 %
15 VDC
2000 mA
86 %
24 VDC
1250 mA
85 %
http://www.tracopower.com
Page 1 of 5



TRACO POWER TMF10105
TMF Series, 5 – 30 W
Input Specifications
Input voltage
– AC Input 90 – 264 VAC
derating of 2%/V below 100 VAC required
– DC Input 120 – 370 VDC
Input frequency 47 – 63 Hz
Leakage current (touch current) < 0.1 mA (at 264 VAC)
Inrush current
5 W models: 10/20 A max. (115/230 VAC)
10 W models: 12/25 A max. (115/230 VAC)
20 W models: 20/40 A max. (115/230 VAC)
30 W models: 30/60 A max. (115/230 VAC)
No-load power consumption < 0.15 W
Output Specifications
Voltage accuracy ± 2 % typ.
Regulation
– Input variation 0.5 % typ.
– Load variation (0–100 %) 1.0 % typ.
Minimum load not required
Hold-up time
5 W models: 20/90 ms min. (115/230 VAC)
10 W models: 12/70 ms min. (115/230 VAC)
20 W models: 10/36 ms min. (115/230 VAC)
30 W models: 14/45 ms min. (115/230 VAC)
Ripple and Noise (20MHz bandwidth) 5 VDC models: 100 mVp-p max.
other models: 1 % of Vout max.
Overload protection (current limit) hiccup, auto recovery
Short circuit current hiccup, indefinite
Output overvoltage protection
zener diode
Max. capacitive load [µF]
Model series
Output
TMF 05 TMF 10 TMF 20 TMF 30
5.0 VDC
8000 13‘000 5000
6800
12 VDC
1200
2600
1500
1600
15 VDC
820
2600
1000
1200
24 VDC 330 600 560 470
General Specifications
Temperature ranges
– operational (natural convection 20 LFM) –25 °C to +70 °C max.
– Storage temperature –40 °C to +85 °C max.
Power derating 5 & 10 Watt models: derating above +55ºC : 4 %/K
20 & 30 Watt models: derating above +50ºC : 3 %/K
Temperature coefficient 0.05 %/K
Cooling convection cooling, no internal fan
Humidity (non condensing) 95 % rel. H max.
Atmospheric pressure 54 kPa – 106 kPa
Altitude during operation 5000 m max.
Switching frequency (at 230 VAC) 66 kHz
Isolation 4000 VAC
MTBF (MIL-HDBK-217F, at 25 °C ground benign)
5 W models: > 450‘000 h
10 W models: > 400‘000 h
20 W models: > 350‘000 h
30 W models: > 300‘000 h
Protection class class II prepared
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
http://www.tracopower.com
Page 2 of 5



TRACO POWER TMF10105
TMF Series, 5 – 30 W
General Specifications (continued)
Safety standards
IEC 60601-1:2005 (3rd Ed.)+ A1:2012,
ANSI/AAMI ES60601-1:2005/(R)2012,
CSA CAN/CSA-C22.2 NO. 60601-1:14
EN 60601-1:2006 + A1:2013 + A12:2014
– Certification documents www.tracopower.com/overview/tmf
Electromagnetic compatibility (EMC), Emissions
– Conducted RI suppression on input EN 55011 class B
– Radiated RI suppression EN 55011 class B
Electromagnetic compatibility (EMC), Immunity
– Electrostatic discharge (ESD) IEC/EN 61000-4-2 4 kV / 8 kV criteria A
– Radiated RF field immunity IEC/EN 61000-4-3 10 V/m criteria A
– Electrical fast transient / burst immunity IEC/EN 61000-4-4 2 kV criteria A
– Surge immunity* IEC/EN 61000-4-5 1 kV criteria A
– Immunity to conducted RF disturbances IEC/EN 61000-4-6 10 V criteria A
– Magnetic field immunity IEC/EN 61000-4-8 30 A/m criteria A
– Mains voltage dips and interruptions IEC/EN 61000-4-11
30% / 10ms criteria B
95% / 5000ms criteria B
Case material plastic (UL94V-0 rated)
Environmental compliance – Reach www.tracopower.com/overview/tmf
– RoHS RoHS directive 2011/65/EU
* For the 5 & 10 W models an external Varistor has to be integreated in the circuit at the converter input (in parallel).