NPC-100 and NPC-120 Series
NovaSensor Disposable Medical Pressure Sensors
Features
Applications
• Solid state, high reliability
• Medical instrumentation
• Media compatibility
• Blood pressure measurement
• High performance
• Infusion pumps
• Factory filled with dielectric gel
• Kidney dialysis machines
• Small size
• Fully tested
• Temperature compensated
• Low cost disposable design
• Designed to AAMI specifications
Amphenol
Advanced Sensors
0.392 in to 0.397 in (9.96 mm to 10.08 mm)
0.393 in to 0.397 in (9.98 mm to 10.08 mm)
NPC-100 and NPC-120 Series Specifications
The NovaSensor NPC-100 and NPC-120 Series pressure sensor are specifically designed for use in disposable medical applications. The devices are compensated and calibrated per the Association for the Advancement of Medical Instrumentation (AAMI) guidelines for industry acceptability. The sensors integrate a high-performance, pressure sensor die with temperature compensation circuitry and gel protection in a small, low-cost package.
The NPC-100 and NPC-120 Series are manufactured in a class 1000 clean-room to minimize possible sources of contamination. A specially designed silicon micromachined sensing element is used to meet or exceed all industry requirements while minimizing assembly and test cost for maximum customer value. Thick-film laser-trimming is employed for final compensation and calibration. Sensitivity is maintained to ±1% and linearity is better than 1% in the physiological operating pressure range.
The NPC-100 and NPC-120 Series are batchmanufactured in ceramic plate form and shipped as an intact array or in Tape and Reel for easy customer automation. This assembly method draws from wellestablished manufacturing techniques used in the electronics industry in order to produce a quality, high volume product.
0.0875 in to 0.0905 in (2.22 mm to 2.30 mm)
0.049 in to 0.052 in (1.24 mm to 1.32 mm)
0.189 in to 0.192 in (4.80 mm to 4.88 mm)
0.236 in to 0.246 in (5.99 mm to 6.25 mm)
Radius 0.005 in (0.13 mm) maximum
0.142 in to 0.145 in (3.61 mm to 3.68 mm)
Material is clear Polycarbonate.
0.282 in to 0.290 in (7.16 mm to 7.37 mm)
NPC-100 Series pressure sensor port interface
0.0875 in to 0.0905 in (2.22 mm to 2.30 mm)
0.049 in to 0.052 in (1.24 mm to 1.32 mm)
0.124 in to 0.126 in (3.14 mm to 3.20 mm)
Radius 0.005 in (0.13 mm) maximum
0.236 in to 0.246 in (5.99 mm to 6.25 mm)
0.142 in to 0.145 in (3.61 mm to 3.68 mm)
Material is clear Polycarbonate.
0.282 in to 0.290 in (7.16 mm to 7.37 mm)
NPC-120 Series pressure sensor port interface
Parameter General Pressure Ranges Overpressure
Value
–50 to 300 125
Units
Notes
mmHg psi
–0.97 psi to 5.8 psi minimum
Electrical @ 72°F (22°C) Unless Otherwise Stated
Input Excitation
1 to 10
VDC
Dielectric Breakdown
10,000
VDC
Risk Current
2
µA
Input Impedance
1800 to 3300
Ω
Output Impedance
285 to 315
Ω
Environmental
Temperature
Compensated
15°C to 40°C
°C
Operating
15°C to 40°C
°C
Storage
–25°C to 70°C °C
Humidity
10 to 90
%
Light Sensitivity
1
mmHg
Operating Product Life 168
hours
Shelf Life 5 years
Mechanical
Weight
<0.0044
lb
Media Interface
Medical grade, dielectric gel
Gel Tube Interface Material
Polycarbonate
Calibrated for 6 VDC 5 Maximum (per AAMI), 5
(59° to 104°F) (59° to 104°F) (–13° to 158°F) maximum (per AAMI BP22)
(<2 g)
Parameter Performance Parameters* Offset Sensitivity Calibration Symmetry Linearity (–30 to 100 mmHg) Linearity (100 to 200 mmHg) Linearity (200 to 300 mmHg) Thermal Coefficient Offset Thermal Coefficient Span Frequency Response Phase Shift Offset Drift
Units
mmHg µV/V/mmHg mmHg % mmHg % output % output mmHg/°C %/°C Hz degrees mmHg/8 hrs
Minimum
–20 4.95 97.5 — — — — — — 1200 — —
Type Maximum Notes
0 20 5 5.05 100 102.5 — ±5 —1 —1 — 1.5 — ±0.3 — ±0.1 —— —5 —1
2
6 6 6 3 3 5 5 4,5
* 1. All values measured at 6 VDC and 71.6°F (22°C) and after five second warm up unless otherwise specified.
2. Output of sensor with no pressure applied and a 150 kW resistor shorted across +IN to +OUT. 3. Over a temperature range of 59°F to 104°F (15°C to 40°C). 4. Normalized offset/bridge voltage—8 hours after 20 seconds. warm-up. 5. Previously qualified, not tested in production. 6. Deviation from straight line drawn through zero and 100 mmHg data points. 7. These sensors are compatible with ETO (ethylene oxide) sterilization. The user is responsible for
establishing and validating suitable process conditions to prevent damage to the sensor.
–
0.160 in to 0.170 in (4.06 mm to 4.32 mm)
Pressure port
Dielectric gel
Vent hole Polycarbonate
gel tube2
Contact pads
0.320 in +0.010/-0.002 (8.13 mm +0.25/-0.05)
1
0.415 in +0.010/-0.002 (10.54 mm +0.25/-0.05)
Gel height +0.001 in/-0.035 in (+0.025 mm/-0.89 mm)
0.109 in to 0.119 in (2.77 mm to 3.02 mm)
1
0.017 in (0.43 mm)
1. Tolerances refle.