Product Data Sheet
4OXV CiTiceL® Oxygen (O2) Gas Sensor Part Number: AAY80-390
Product Datasheet 4OXV Oxygen CiTiceL®
...
Product Data Sheet
4OXV CiTiceL® Oxygen (O2) Gas Sensor Part Number: AAY80-390
Product Datasheet 4OXV Oxygen CiTiceL®
Document Purpose
The purpose of this document is to present the performance specification of the 4OXV oxygen sensor.
This document should be used in conjunction with the 4OXV Characterisation Note, Operating Principles (OP02) and the Product Safety Datasheet (PSDS 4).
The data provided in this document are valid at 20°C, 50% RH and 1013 mBar for 3 months from the date of sensor manufacture. For guidance on sensor performance outside of these limits, please refer to the 4OXV Characterisation Note.
Output signal can drift below the lower limit over time. For guidance on the safe use of the sensor, please refer to the Operating Principles OP02.
Doc Ref : 4OXV.indd ECN I 4473 Issue 6 13th July 2016
Page 1 of 3
Web: www.citytech.com Email
[email protected] Call +44(0) 23 9228 8100
The Right Sensor Can Save A Life
Product Data Sheet
4OXV CiTiceL® Oxygen (O2) Gas Sensor Part Number: AAY80-390
Key Features and Benefits
False alarm immunity Enhanced response time in extreme applications Reliably meets stated life Superior environmental performance
Technical Specifications
MEASUREMENT Technology Electrochemical Measurement Range 1-25% vol. O2 Maximum Overload 30% vol. O2 Output Signal* 0.10 ± 0.02 mA in Air Response Time (T90)* <15 Seconds Zero Current (Offset)* <0.6% vol. O2
Linearity Can be considered linear in many cases. See Operating Principles...