AC-DC / External Freestanding Medical Adapter
Embedded Power for Business-Critical Continuity
DPT50-M Series
50 Watts
Total Power: Input Voltage: # of Outputs:...
Description
Embedded Power for Business-Critical Continuity
DPT50-M Series
50 Watts
Total Power: Input Voltage: # of Outputs: 30 - 50 Watts 127 - 300 VDC 90 - 264 VAC Triple
Rev. 10.02.08_127 DPT50-M Series of 2
RoHS
Electrical Specifications
Special Features
Medical safety approvals CE Mark EMC & LVD Universal AC input Fully regulated output Overpower and short circuit protection Constant voltage High efficiency High MTBF IEC320 AC receptacle, 3 pin (type C14) Built in EMI filter (CISPR 22 Class B) LED power good indicator Input power <74 watts Complies with EN61000-3-2
Input
Input range Frequency Inrush current Efficiency EMI RFI Safety ground leakage current 90-264 VAC (wide range) 127-300 VDC 47-440 Hz <60 A peak @ 230 VAC, cold start @ 25°C 80% @ typical full load, @115VAC FCC Part 15, Subpart B Class B & EN55022 (CISPR22) Class B 275uA @ 50/60 Hz, 264 VAC input
Output
Maximum power 50 W (DPT51-M, 30W) Hold-up time 10 ms typical at full load @ 115VAC Overpower protection Output short circuit protection auto recover. Overpower protected @ 110 - 160% above maximum rating Overvoltage protection 20-45% above nominal output maximum Caable/2 meter DC cable with 5 pin Din connector
Environmental Specifications
Operating temperature: 0°C to +40°C ambient. Derate each output as 2.5% per degree from 40° to 60°C. -20°C start up. Storage temperature: -40°C to +85°C Electromagnetic susceptibility: Designed to meet EN61000-4; -2, -3, -4, -5, -6, -8,...