DatasheetsPDF.com

Medical Adapter. DPT53-M Datasheet

DatasheetsPDF.com

Medical Adapter. DPT53-M Datasheet






DPT53-M Adapter. Datasheet pdf. Equivalent




DPT53-M Adapter. Datasheet pdf. Equivalent





Part

DPT53-M

Description

AC-DC / External Freestanding Medical Adapter



Feature


Embedded Power for Business-Critical Con tinuity DPT50-M Series 50 Watts Total Power: Input Voltage: # of Output s: 30 - 50 Watts 127 - 300 VDC 90 - 26 4 VAC Triple Rev. 10.02.08_127 DPT50-M Series  of 2 RoHS Electrical Specif ications Special Features • Medical s afety approvals • CE Mark EMC & LVD Universal AC input • Fully regulat ed output • Overpower and .
Manufacture

Emerson

Datasheet
Download DPT53-M Datasheet


Emerson DPT53-M

DPT53-M; short circuit protection • Constant vo ltage • High efficiency • High MTBF • IEC320 AC receptacle, 3 pin (type C14) • Built in EMI filter (CISPR 22 Class B) • LED power good indicator Input power <74 watts • Complies w ith EN61000-3-2 Input Input ra­nge Fr equency Inrush current Efficiency EMI R FI Safety ground leakage current 90-264 VAC (wide range) 127-300 VDC 47-44.


Emerson DPT53-M

0 Hz <60 A peak @ 230 VAC, cold start @ 25°C 80% @ typical full load, @115VAC FCC Part 15, Subpart B Class B & EN5502 2 (CISPR22) Class B 275uA @ 50/60 Hz, 2 64 VAC input Output Maximum power 50 W (DPT51-M, 30W) Hold-up time 10 ms typi cal at full load @ 115VAC Overpower pro tection Output short circuit protection auto recover. Overpower protected @ 11 0 - 160% above maxi.


Emerson DPT53-M

mum rating Overvoltage protection 20-45% above nominal output maximum Caable/2 meter DC cable with 5 pin Din connector Environmental Specifications Operatin g temperature: 0°C to +40°C ambient. D­­erate each output as 2.5% per degr ee from 40° to 60°C. -20°C start up. Storage temperature: -40°C to +85°C Electromagnetic susceptibility: Designe d to meet EN61000-4; ­­­-2,.

Part

DPT53-M

Description

AC-DC / External Freestanding Medical Adapter



Feature


Embedded Power for Business-Critical Con tinuity DPT50-M Series 50 Watts Total Power: Input Voltage: # of Output s: 30 - 50 Watts 127 - 300 VDC 90 - 26 4 VAC Triple Rev. 10.02.08_127 DPT50-M Series  of 2 RoHS Electrical Specif ications Special Features • Medical s afety approvals • CE Mark EMC & LVD Universal AC input • Fully regulat ed output • Overpower and .
Manufacture

Emerson

Datasheet
Download DPT53-M Datasheet




 DPT53-M
Embedded Power for
Business-Critical Continuity
DPT50-M Series
50 Watts
Total Power:
Input Voltage:
# of Outputs:
30 - 50 Watts
127 - 300 VDC
90 - 264 VAC
Triple
Rev. 10.02.08_127
DPT50-M Series
 of 2
RoHS
Special Features
• Medical safety approvals
• CE Mark EMC & LVD
• Universal AC input
• Fully regulated output
• Overpower and short
circuit protection
• Constant voltage
• High efficiency
• High MTBF
• IEC320 AC receptacle,
3 pin (type C14)
• Built in EMI filter
(CISPR 22 Class B)
• LED power good indicator
• Input power <74 watts
• Complies with EN61000-3-2
Safety
UL UL60601-1
CSA CSA-C22.2 NO. 601.1-M90
VDE 07501/EN60601-1 (IEC601)
CE EMC & LVD
Electrical Specifications
Input
Input ra­nge
90-264 VAC (wide range) 127-300 VDC
Frequency
47-440 Hz
Inrush current
<60 A peak @ 230 VAC, cold start @ 25°C
Efficiency
80% @ typical full load, @115VAC
EMI RFI
Safety ground
leakage current
FCC Part 15, Subpart B Class B & EN55022 (CISPR22) Class B
275uA @ 50/60 Hz, 264 VAC input
Output
Maximum power
50 W (DPT51-M, 30W)
Hold-up time
10 ms typical at full load @ 115VAC
Overpower protection Output short circuit protection auto recover.
Overpower protected @ 110 - 160% above maximum rating
Overvoltage protection 20-45% above nominal output maximum
Caable/2 meter
connector
DC cable with 5 pin Din
Environmental Specifications
Operating temperature: 0°C to +40°C ambient. D­­erate each output as 2.5% per degree
from 40° to 60°C. -20°C start up.
Storage temperature: -40°C to +85°C
Electromagnetic
susceptibility:
Designed to meet EN61000-4; ­­-2, -3, -4, -5, -6, -8, -11 Level 3
Humidity:
Operating; non-condensing 10% to 95% RH
Vibration:
IEC68-2-6 to the levels of IEC721-3-2
MTBF demonstrated: >550,000 hours at full load and 25°C ambient conditions




 DPT53-M
Embedded Power for
Business-Critical Continuity
Ordering Information
Model
Number
DPT51-M
DPT52-M
DPT53-M
DPT54-M
Output
Voltage
+3.3V
+5V
+12V
+5V
+12V
-12V
+5V
+15V
-15V
+5V
+24V
+12V
Minimum
Load
0.8A
0.1A
0A
0.5A
0.1A
0A
0.5A
0.1A
0A
0.5A
0.1A
0A
Maximum Load Peak Load 1 Regulation 2
9A
3A
0.5A
8A
3A
0.5A
8A
2.4A
0.5A
8A
1.5A
0.5A
9A
4A
1A
9A
4A
1.0A
9A
3.2A
0.7A
9A
2A
0.7A
±4%
±6%
±5%
±2%
±5%
±5%
±2%
±5%
±5%
±2%
±7%
±5%
Ripple P/P (PARD)3
50 mV
50mV
120mV
50 mV
120mV
120mV
50 mV
150mV
150mV
50 mV
240mV
120mV
1. Peak current lasting <15 seconds with a maximum 10% duty cycle.
2. At 25°C including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings.
3. Peak-to-peak with 20 MHz bandwidth and 10 µF (tantalum capacitor) in parallel with a 0.1 µF capacitor at rated
line voltage and load ranges.
Mechanical Drawing
NOTES:
1. Specifications subject to change without notice.
2. All dimensions in inches (mm), tolerance is ± 0.02” ± (0.5mm)
3. Warranty: 2 year
4. Weight: 1.44 lb./ 0.65 kg
5. AC input power cord sold separately.
6. Specifications are for convection rating at factory settings
at 115VAC input, 25°C unless otherwise stated
COM
5
2V1 4 V3
V2 1V1
3 Shield
COM
PINOUT ARRANGEMENT
Rev. 10.02.08_127
­Americas
DPT50-M Series
 of 2
5810 Van
Allen Way
Carlsbad, CA 92008
USA
Telephone: +1 760 930 4600
Facsimile: +1 760 930 0698
Europe (UK)
Waterfront Business Park
Merry Hill, Dudley
West Midlands, DY5 1LX
United Kingdom
Telephone: +44 (0) 1384 842 211
Facsimile: +44 (0) 1384 843 355
Asia (HK)
14/F, Lu Plaza
2 Wing Yip Street
Kwun Tong, Kowloon
Hong Kong
Telephone: +852 2176 3333
Facsimile: +852 2176 3888
For global contact, visit:
www.powerconversion.com
techsupport.embeddedpower
@emerson.com
While every precaution has been taken to ensure
accuracy and completeness in this literature, Emerson
Network Power assumes no responsibility, and disclaims
all liability for damages resulting from use of this
information or for any errors or omissions.
Emerson Network Power.
The global leader in enabling
business-critical continuity.
AC Power
Connectivity
DC Power
Embedded Computing
Embedded Power
Monitoring
Outside Plant
Power Switching & Controls
Precision Cooling
Racks & Integrated Cabinets
Services
Surge Protection
EmersonNetworkPower.com
Emerson Network Power and the Emerson
Network Power logo are trademarks and
service marks of Emerson Electric Co.
©2008 Emerson Electric Co.










Recommended third-party DPT53-M Datasheet






@ 2014 :: Datasheetspdf.com :: Semiconductors datasheet search & download site (Privacy Policy & Contact)