131 Business Center Drive Suite A‐3, Ormond Beach, FL 32174
TEL: 877.678.4259 | FAX: 866.422.2356 EMAIL: sales@co2me...
131 Business Center Drive Suite A‐3, Ormond Beach, FL 32174
TEL: 877.678.4259 | FAX: 866.422.2356 EMAIL:
[email protected] | www.co2meter.com
TR250Z Manual
Manual
Revision 1.0 – March 2012
1
131 Business Center Drive Suite A‐3, Ormond Beach, FL 32174
TEL: 877.678.4259 | FAX: 866.422.2356 EMAIL:
[email protected] | www.co2meter.com
Description
Display – Main interface to view current sensor readings and menu state. Will blink “STRT” upon applying power while the sensor warms up and stabilizes Buttons 1, 2, 3 – Allow for entering the menu, adjusting values, and performing a calibration. S LED – Status LED. Blinks every time a complete measurement cycle is performed and the value is updated. Serves as a heartbeat for transmitter operation. E LED – Error LED. Will light if any abnormal conditions are detected within the device, primarily the sensor falling out of its recommended operating specifications will trigger this lamp. P LED – Power LED. Will light continuously when power is applied to unit. The OX100 Device has three main navigation buttons. These buttons allow you to access all aspects of sensors operation and test or calibrate different functional modules.
Standard Operation
The O2 transmitter comes factory calibrated for easy installation. During its life cycle calibration may become periodically necessary. We recommend performing a span O2 calibration every month for guaranteed operation and ensured acc...