DatasheetsPDF.com

VOC Monitor. MS1100 Datasheet

DatasheetsPDF.com

VOC Monitor. MS1100 Datasheet






MS1100 Monitor. Datasheet pdf. Equivalent




MS1100 Monitor. Datasheet pdf. Equivalent





Part

MS1100

Description

VOC Monitor



Feature


MS1100 VOC Monitor V1.0 March 2011 Prod uct Description The MS1100 is a Volatil e Organic Compound (Hydrocarbon) monito r designed as an alarm system for clean and waste water applications. It provi des on-line, real time monitoring and a larm generation for the protection of e quipment and the environment. The instr ument samples air from the headspace in a sampling tank f.
Manufacture

multisensor

Datasheet
Download MS1100 Datasheet


multisensor MS1100

MS1100; or high sensitivity applications such as clean water intakes, from drains throu gh the use of a sampling fan or directl y from the air. The MS1100 provides ver y high sensitivity, to ppb levels, fast processing and low cost of ownership t hrough long servicing and validation pe riods. The instrument can be powered f rom either AC or DC supplies, provides various communicat.


multisensor MS1100

ions options and is available with an op tional sampling system. Summary Specif ication Conforms to UL 61010-1 / EN 610 10-1:2010 EN61326-1:2006 Absolute Accu racy For repeated measurement of 200ppb Toluene in Water: ±5% Dimensions Hei ght 400mm (15.75ins) x Width 400mm (15. 75ins) x Depth 200mm (7.9ins) Analysis time 60s, repetition rate 5, 10, 15 mi ns Dynamic Range 1.


multisensor MS1100

– 1,000 ppb in water Applications Pr otection of clean water intakes to wate r treatment works Detection of hydrocar bons in waste water drains Detection of discharges from industrial facilities Monitoring of the environment Precisio n User programmable, typical levels: Al arm1 = 15% - 30% ±2 Alarm2 = 25% - 60% ±5 Repeatability For repeated measur ement of 200ppb Toluen.

Part

MS1100

Description

VOC Monitor



Feature


MS1100 VOC Monitor V1.0 March 2011 Prod uct Description The MS1100 is a Volatil e Organic Compound (Hydrocarbon) monito r designed as an alarm system for clean and waste water applications. It provi des on-line, real time monitoring and a larm generation for the protection of e quipment and the environment. The instr ument samples air from the headspace in a sampling tank f.
Manufacture

multisensor

Datasheet
Download MS1100 Datasheet




 MS1100
MS1100 VOC Monitor
V1.0 March 2011
Product Description
The MS1100 is a Volatile Organic Compound
(Hydrocarbon) monitor designed as an alarm
system for clean and waste water applications. It
provides on-line, real time monitoring and alarm
generation for the protection of equipment and
the environment.
The instrument samples air from the headspace in
a sampling tank for high sensitivity applications
such as clean water intakes, from drains through
the use of a sampling fan or directly from the air.
The MS1100 provides very high sensitivity, to ppb
levels, fast processing and low cost of ownership
through long servicing and validation periods.
The instrument can be powered from either
AC or DC supplies, provides various
communications options and is available with
an optional sampling system.
Summary Specification
Conforms to
UL 61010-1 / EN 61010-1:2010
EN61326-1:2006
Dimensions
Height 400mm (15.75ins) x Width 400mm
(15.75ins) x Depth 200mm (7.9ins)
Dynamic Range
1 1,000 ppb in water
Precision
User programmable, typical levels:
Alarm1 = 15% - 30% ±2
Alarm2 = 25% - 60% ±5
Repeatability
For repeated measurement of 200ppb
Toluene in Water: ±2%
Absolute Accuracy
For repeated measurement of 200ppb
Toluene in Water: ±5%
Analysis time
60s, repetition rate 5, 10, 15 mins
Applications
Protection of clean water intakes to water
treatment works
Detection of hydrocarbons in waste water drains
Detection of discharges from industrial facilities
Monitoring of the environment
Version 1.0
1
Confidential
http://www.Datasheet4U.com




 MS1100
Enclosure
Dimensions
Height 400mm (15.75ins) x Width 400mm (15.75ins) x Depth 200mm (7.9ins)
Supplied with wall mounting kit
Material
Steel
Weight
12kg (27lbs)
IP Rating
IP65 (IP66 option available)
Connections
Sample inlet: 1/4” 28 UNF screw fitting to 1/8” PTFE tube
Purge inlet:
1/4” 28 UNF screw fitting to 1/8” PTFE tube
Exhaust output: 1/4” 28 UNF screw fitting to 1/8” PTFE Tube
Sample Presentation System (Optional extra)
Volume:
3 10L
Flow:
1l/min
Headspace:
Surface area:
1l 2l
500 1000 cm2
MS1100
Purge
Exhaust
Gas under test
Sample Water
Figure 1: Typical Setup
Version 1.0
2
Confidential




 MS1100
Operating Conditions
Parameter
AC Power Voltage
DC Power
Power Consumption
Temperature
Relative Humidity
Sample Gas Flow
Sample Temperature
Notes
50/60Hz Systems, alternative to DC
Alternative to AC
Continuous
Min Max Units
85 264 V rms
19 36
V
50 W
0 40 C
0 85 %
40 60 ml/s
2 40 C
Technical Specification
Safety
EMC
UL Std 61010-1
EN61326-1:2006
EN 61010-1:2010
Modes of operation
Normal operation:
5, 10 or 15 minute sampling
Raw:
Measurement without historical data being used
Test:
Continuous sampling every 2 seconds vs single purge measurement
Performance (Tambient = 20C, Tsample = 18C, see setup in Fig 2, Sensor Gain = 1, Toluene in distilled water)
Sensitivity
1 ppb toluene
Dynamic Range
1 1,000 ppb in water
Precision
Alarm level user programmable, typical levels:
Alarm1 = 15% - 30% ±2
Alarm2 = 25% - 60% ±5
Repeatability
For repeated measurement of 200ppb Toluene in Water: ±2%
Absolute Accuracy
For measurement of 200ppb Toluene in Water: ±5%
Analysis time
60s, repetition rate 5, 10, 15 mins
Measurement Stability
Maximum time between validation tests: 180 days
Version 1.0
3
Confidential



Recommended third-party MS1100 Datasheet






@ 2014 :: Datasheetspdf.com :: Semiconductors datasheet search & download site (Privacy Policy & Contact)