DatasheetsPDF.com

VPM240-4170

TRIAD
Part Number VPM240-4170
Manufacturer TRIAD
Description POWER TRANSFORMER
Published Oct 2, 2019
Detailed Description POWER TRANSFORMER CHASSIS MOUNT : TOROIDAL MEDICAL SERIES VPM240-4170 Description: The toroidal construction inherentl...
Datasheet PDF File VPM240-4170 PDF File

VPM240-4170
VPM240-4170


Overview
POWER TRANSFORMER CHASSIS MOUNT : TOROIDAL MEDICAL SERIES VPM240-4170 Description: The toroidal construction inherently reduces stray fields, increases efficiency and minimizes size compared to traditional EI transformers.
The addition of a Flux Band further reduces the remaining stray fields.
The shield between Primary and Secondary improves safety, reduces common mode signals and minimizes leakage current.
Built with a Class F (155°) insulation system.
A 140°C self-resetting thermal switch is included in each primary.
Electrical Specifications (@25C) 1.
Maximum Power: 1000VA 2.
Input Voltages: 100, 120, 220, 240VAC, 50/60Hz 3.
Output Voltages: 120VAC @8.
34A or 240VAC CT @ 4.
17A 4.
Voltage Regulation: 3.
0% TYP from full load to no load 5.
Temperature Rise: 50°C TYP 6.
Hipot: 4000VAC, Primary to Secondary, Primary & Secondary to Shield & mounting surface 7.
Efficiency: 96% TYP.
@ full load 8.
Earth Leakage: ≤10µA (See Fig.
1), Patient Leakage: ≤ 100µA (see Fig.
2) Agency File: UL: File E122529, UL 60601-1/(R) 2012 Medical Electrical Equipment – Part 1 with 2 MOPP CE: ES 60601-1 (IEC 60601-1:2005, MOD) cUL: C22.
2 No.
60601-1:14, Medical Electrical Equipment – Part 1 CB Certified.
RoHS 2011/65/EU Photo for illustration only OD Dimensions: Inches (mm) O.
D.
I.
D.
HT.
* 7.
2 (183) 2.
3(59) 2.
9(74) *Add 0.
188 (3) to the height for mounting hardware Weight: 8.
5Kg Mounting: Transformer is provided with one metal mounting plate, two rubber pads, M8 x 80mm bolt, nut, spring and flat washer.
MOUNTING METAL PLATE BOLT FLAT WASHER RUBBER PAD Connections: Transformer is provided with 12” (305mm) long, 0.
5” (12.
7mm) stripped and tinned, stranded UL 1015 lead wire.
Primaries are 18AWG, Secondaries are 16AWG, and Shield is 20AWG.
The GRN/YEL shield lead is typically grounded.
Do not lift transformer by leads! H SPRING WASHER RUBBER NUT PAD Input Options: 100VAC: Input to Gray & Blue, jumper White & Brown, jumper Blue & Violet.