DatasheetsPDF.com

LPS54-M

Emerson
Part Number LPS54-M
Manufacturer Emerson
Description AC-DC / 10W - 85W Medical
Published Oct 25, 2009
Detailed Description Embedded Power for Business-Critical Continuity LPS50-M Series 60 Watts Total Power: Input Voltage: # of Outputs:...
Datasheet PDF File LPS54-M PDF File

LPS54-M
LPS54-M



Overview
Embedded Power for Business-Critical Continuity LPS50-M Series 60 Watts Total Power: Input Voltage: # of Outputs: 55 - 60 Watts 90 - 264 VAC 127 - 300 VDC Single Rev.
1.
08.
09_149 LPS50-M Series  of 3 Special Features • Medical Safety Approvals • Universal input • Less than 1U high • 2” x 4” footprint • Remote sense • Overload and short circuit protection • Adjustable output voltage • High efficiency • High MTBF • Built in EMI filter (CISPR 22 Class B) • LED power good indicator • Input power < 74 watts • Complies with EN61000 -3-2 • UL Class I approved • LPX50 enclosure kit available • Dual AC fuses Electrical Specifications Input Input ra­nge: Frequency: Inrush current: Input power: Efficiency: EMI/RFI: Safety ground leakage current: 90 - 264 VAC (wide range) 127-300 Vdc 47-440 Hz < 60 A peak @ 230 VAC, cold start @ 25°C < 74 Watts 80% - 85% typical at full load FCC Class B conducted; CISPR 22 Class B conducted; EN55022 Class B conducted, VDE0878PT3 Class B conducted 275uA @ 50/60 Hz; 264 VAC input Output Maximum power: Adjustment range: Hold-up time: Overload protection: 60W for convection (LPS52, 55 W) ±20% minimum 10/20 ms 115/230 VAC input line Short circuit protection on all outputs.
Case overload protected @ 110-160% of normal rating Overvoltage protection: 30-50% above nominal output Safety • UL: • CSA: • VDE: • CE: UL 60601-1 CSA-C22.
2 No.
601.
1-M90 EN60601-1 LVD & EMC Remote sense: Compensatesd for 0.
5 V lead drop max.
Will operate without remote sense connected.
Reverse connection protected.
Embedded Power for Business-Critical Continuity Environmental Specifications Operating temperature: Storage temperature: Electromagnetic susceptibility: Humidity: Vibration: MTBF demonstrated: Rev.
1.
08.
09_149 LPS50-M Series  of 3 0° to 50 °C ambient derate each output as 2.
5% per degree from 50° to 70 °C.
-20 °C start up -40 °C to +85 °C Designed to meet EN61000-4; -2, -3, -4, -5, -6, -8, -11 Level 3 Operating; non-condensing 10% to 95% RH ...