DatasheetsPDF.com

JHM1005D15

XP Power
Part Number JHM1005D15
Manufacturer XP Power
Description DC to DC Converter and Switching Regulator
Published Dec 2, 2011
Detailed Description www.DataSheet.co.kr DC DC 10 Watts JHM10 Series • • • • • • • xppower.com International Medical Approvals 4000 VAC R...
Datasheet PDF File JHM1005D15 PDF File

JHM1005D15
JHM1005D15


Overview
www.
DataSheet.
co.
kr DC DC 10 Watts JHM10 Series • • • • • • • xppower.
com International Medical Approvals 4000 VAC Reinforced Insulation Meets IEC60601 1, 3rd Edition 2 µA Patient Leakage Current DIP 24 Package EN55011 Level A With No External Components 3 Year Warranty Specification Input Input Voltage Range • 5 V (4.
5-9 VDC) 12 V (9-18 VDC) 24 V (18-36 VDC) • See table • 20 A max at 36 V • Pi network • 2 µA max • 5 V models, on at >4.
4 V, off <4.
2 V 12 V models, on at >8.
8 V, off <8.
3 V 24 V models, on at >17.
5 V, off <17.
0 V • 5 V models 15 V for 3 s 12 V models 25 V for 3 s 24 V models 50 V for 3 s General Efficiency Isolation • See tables • 4000 VAC for 1 min.
double/reinforced with a working voltage of 250 VAC.
Meets 2 x MOPP per 3rd edition of IEC60601-1 5000 VAC for 10 ms in accordance with IEC60664-1 • 20 pF max • 80 kHz to 1.
2 MHz variable • 20.
0 W/in3 • >1 MHrs typical to MIL-STD-217F at 25 °C, GB Input Current Inrush Current Input Filter Patient Leakage Current Undervoltage Lockout Input to Output Capacitance Switching Frequency Power Density MTBF Input Surge Environmental Operating Temperature Case Temperature Storage Temperature Operating Humidity Cooling • • • • • -40 °C to +80 °C, see derating curve +100 °C max -40 °C to +100 °C 5-90%, non-condensing Natural convection Output Output Voltage Output Voltage Trim Minimum Load Initial Set Accuracy Start Up Delay Start Up Rise Time Line Regulation Load Regulation Cross Regulation • • • • • • • • • See table ±10% No minimum load required ±1% max on V1, ±2% max on V2 5 ms typical 2 ms typical ±0.
3% max ±2% max 0% to 10% load, ±1% max 10% to 100% load ±4% max on dual with one output set to 50% load and the other varied from 10% to 100% load (D05 is 20% to 100%) 4% max deviation, recovery to within 1% in <500 µs for a 50% load change at 0.
25 A/µs rate 1% pk-pk max at 20 MHz bandwidth 120% - 200%, trip and restart 115% - 140% ±0.
03/°C max EMC & Safety Emissions Immunity ESD Immunity Radiated Immuni...



Similar Datasheet