DatasheetsPDF.com

SMM201-12-3.3

Conversion Devices
Part Number SMM201-12-3.3
Manufacturer Conversion Devices
Description Medical Single or Dual Output
Published May 13, 2014
Detailed Description ADVANCED DATA Medical Single or Dual Output 3.0" W x 5.0" L x 1.38" H EAGLE SERIES Model SMM201 Total Output Power Outp...
Datasheet PDF File SMM201-12-3.3 PDF File

SMM201-12-3.3
SMM201-12-3.3


Overview
ADVANCED DATA Medical Single or Dual Output 3.
0" W x 5.
0" L x 1.
38" H EAGLE SERIES Model SMM201 Total Output Power Output Channels Input Mains Voltage 150/200 watt Single or Dual 90-264VAC FEATURES • • • • 200 watt single or dual output ZVS power supply High Efficiency 90% typical Open PCB construction Output voltages: V1 Hi-current Main output available with any voltage from 12 to 60 V up to 16.
7A current rating V2 Auxiliary output available with any voltage from 3.
3 to 12.
0 V rated up to 5.
0A maximum Input GENERAL SPECIFICATIONS Input voltage ……………….
.
.
Universal input 90 to 264 VAC Power Factor ……………….
.
0.
95 Frequency .
.
.
47-63 Hz Input current………………….
2.
2A @ 115 VAC 1.
2A @ 230 VAC Inrush current .
.
.
35A @ 115 VAC max.
, cold start @ 25° C 70A @ 230 VAC Efficiency 90% typical at full load, nom.
line EMI filter .
FCC Class “B”, CISPR 22 Class “B” and EN55022 Class B.
Earth leakage current .
.
…….
.
less than 135 µA @ 132 VAC input.
Less than 270µA @ 264 VAC input.
Lower leakage models optional.
Contact factory for option code Isolation voltage: Input to output ….
.
.
4200 VAC Input to chassis ….
1500 VAC • • • • • 150 watt convection rating Hi-density Compact footprint of 3.
0”W x 5.
0”L x 1.
38”H UL2601, IEC601 and CSA22.
2 No.
601 pending Low leakage current less than 300 µA @ hi-line 264 VAC Dual AC fusing on both lines Output Output voltage …………….
…See Table Output voltage trim ………….
±1 V typical Maximum power .
.
.
200 watts Minimum Load ……………….
No minimum load required Start-up delay ……………….
.
2 s max.
Start-up rise time ………….
.
.
.
2 ms Hold-up time.
>12 ms @ 150 W load, 115 VAC nominal line at factory voltage setting.
Line Regulation ………….
….
.
1% max.
, 0.
5% typical Load Regulation ……….
.
….
.
.
3% max.
, 1% typical Transient Response ……….
.
.
4% max.
deviation, 500 µs recovery time for a 50% load change Ripple & Noise …………….
.
.
.
1% pk-to-pk expressed a...